FLUKE Impulse 7000DP Defibrillator Analyzer and Pacemaker Tester
Do you want to purchase this product in bulk quantity? Fill the form and our team will send you the best offer.
The Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer ensures proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 7000DP is superior in compliance, technology, efficiency and ease of use. A sleek and rugged exterior is paired with first-class technology, putting this defib tester in a league of its own. Standardize test procedures, capture, and print results with Ansur test-automation software, or use the Impulse 7010 Defibrillator Selectable Load Accessory to ensure IEC 60601-2-4 and AAMI DF80 compliance.
Key Features :-
- Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω,100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional)
- Lown, Edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility
- AED technology compatibility
- First-class measurement accuracy ± 1 % of reading + 0.1 J
- Intuitive user interface and backlight, easy-to-read display
- Portable, rugged, easy to carry
- Long-lasting, rechargeable battery
- Pacer input protected against defibrillator output (7000DP only)
- 10 independent ECG outputs that provide 12 lead combinations for standardized clinical signals
- Flexible heart-rate settings (1 BPM step) facilitate rate meter accuracy and alarm testing
- DSP-based measurements enable future firmware and waveforms upgrade
- Unique integrated posts for secure connections
- Two-year extended warranty (no-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center)
- Optional Ansur test automation software to standardize testing procedures, capture waveforms and test results, and print and document test results
- Designed, tested, and built to incomparable Fluke quality standards
General specifications :-
Temperature Operating | 10 °C to 40 °C (50 °F to 104 °F) |
Storage | -20 °C to +60 °C (-4 °F to +140 °F) |
Humidity | 10 % to 90 % non-condensing |
Display | LCD display |
Communications | USB device port for computer control |
Modes of Operation | Manual and remote |
Power | Internal rechargeable NiMH battery pack for nine hours (typical) operation after full charge or the battery charger can operate the analyzer and charge the battery simultaneously |
Battery Charger | 100 V to 240 V input, 15 V/1.5 V A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle |
Enclosure | ABS plastic housing |
Dimensions | 32 cm x 24 cm x 13 cm (13 in x 9.5 in x 5 in) |
Weight | 3.02 kg (6.6 lb, 0.1 oz) |
Safety Standards | *CE: IEC/EN61010-1 2nd Edition *Pollution degree 2 CAN/CSA-C22.2 No. 61010-1 *UL61010-1 C-Tick: Australian EMC |
No reviews found