B Braun HNS 12 Peripheral Nerve Stimulator with Mapper
Do you want to purchase this product in bulk quantity? Fill the form and our team will send you the best offer.
Stimuplex® HNS12 with SENSe is a peripheral nerve stimulator for single and continuous nerve block and was rated the best overall stimulator in a clinical evaluation1 of thirteen machines from five different manufacturers, scoring a perfect 20/20.
Design Advantages
- A patented mode of functionality called SENSe, (Sequential Electrical Nerve Stimulation) can be turned on to help find the nerve sooner and maintain the motor response for longer.
- Pulse duration can be changed, (0.1, 0.3 and 1.0 Ms) to target motor/ sensory nerve components for greater flexibility.
- Disconnect and current threshold alarms provide audible alerts to reduce patient risk.
- Resistance monitoring shows the relationship between actual patient and desired current level for increased safety.
Safety Data:
Latex Information - Not Made With Natural Rubber
Storage and Shipping:
Total shelf life (Months)- 999
Ordering Unit - ST
Sterility:
Sterile - No
Type: Stimuplex® HNS 12
Technical Features
- Instrument type: BF Battery: 9V (alkaline)
- Power consumption: 6 mA (8 mA max.)
- Stimulation current: Î = 5 mA (max.) (0–12 kΩ)
- Stimulation voltage: Û = 95 V (max.)
- Stimulation frequency: 1 Hz / 2 Hz ±1%/SENSe (3 Hz)
- Stimulus duration: 0.05 ms – 0.10 ms – 0.30 ms – 0.50 ms – 1.00 ms ±1% SENSe (0.10 ms - 0.10 ms - 0.15 ms to 1.0 ms)
- Allowable load impedance: 0 kΩ–12 kΩ
- Current measuring accuracy: ± 0.02 mA
- Impedance measuring range: 1 kΩ – 90 kΩ for target stimulation current > 0.5 mA
- Impedance measuring accuracy: ±10%/±20% for target stimulation current>1mA/<= 1mA
- Sound pressure level: 51 dB / 54 dB / 63 dB for stimulation/warning/error
- Weight: 250g
- Operational environmental conditions: 0 – 50° C, max. 90 % relative humidity, no condensation